Set of computer – moisture analyzer commands – RADWAG MAC 210 Moisture Analyzer User Manual
Page 56

56
17.
SET OF COMPUTER – MOISTURE ANALYZER COMMANDS
Function
INTERFACE RESET
Command
R CR LF
(zeroing of currently carried out commands, i.e. “tare”
which waits for stable result).
 
Function 
SEND ALL IMPLEMENTED COMMANDS FROM THE 
INSTRUMENT 
Command
PC CR LF
(sends information on all implemented commands in
the program from the instrument)
 
Function 
SEND STABLE RESULT IN BASIC UNIT
Command
S CR LF
(sends result in basic unit from the instrument on stable
mass readout)
Function
IMMEDIATELY SEND THE RESULT IN BASIC UNIT
Command
SI CR LF
(sends result in basic unit from the instrument)
 
 
Function 
SEND THE RESULT IN CURRENT UNIT
Command
SU CR LF
(sends the result in current used result on stable mass
readout)