Sample mass control – RADWAG MAC 210 Moisture Analyzer User Manual

Page 51

background image

51

In order to exit the option of using drying program libraries, disable access to

libraries in moisture analyzer’s menu (see point 9.3. of this user manual).

Changing settings of any drying program (auto switch off mode, drying

mode, drying temperature, data to be displayed and printed) in the library,

requires selecting one of drying programs, and then acting as specified in
point 14:

DRYING PROCESS WITH USE OF SHORTENED MENU.

Additionally, each drying procedure stored in the library allows for using an option of

sample mass control. When enabling the option, a moisture analyzer software

informs the user whether the mass to be tested is within set thresholds, regarding

the mass and tolerance values, on condition that the parameters were determined

for the sample. The sample mass control is visualized by displaying pictograms LO,

OK or HI on moisture analyzer display. The option is implemented for information

purpose only and facilitates a user in preparing the same or very similar sample to

be tested, thus increasing repeatability of measurements. If the pre-determined

mass control criteria are not met (i.e. pictograms LO or HI are displayed), the

moisture analyzer still starts the drying process. Settings of mass control option are

provided below.

15.1.

Sample mass control

Sample mass control option is active only in case of enabled access to library

of drying programs – P6-01 Libr (parameter set to YES).

The mass control value and tolerance value are set in the following way: