If moisture balance software detects an error a proper error message appears on

the graphic display. The error message contains number of error which informs

about its source.

Error message

Error description

Er1 Hi

Start mass exceeds permissible range (it refers to start

and adjustment – start mass)

Er2 nu

Low limit of converter range exceeded

Er3 Fu

High limit of converter range exceeded

Er4 ro

Value of mass or temperature exceeds permissible range

(it refers to mass or temperature adjustment – final mass

or temperature)

Er5 Ad

A/D converter error

Er6 to

Operation time exceeded (tarring or printing caused by

lack of stability status in set time)

Er7 ou

Inserted value is out of range (reg. setting reference

mass, date, etc.)

19.

PRINCIPLES OF USE

For measurement temperatures ranging from 161°C to 250°C time interval for

temperature meintenance is estimated proportionally as follows, 60 min for 161°C –

20 min for 250°C.
For drying process in temperature 250°C, Max temp is maintained for 20 min, next

the program automatically lowers the temperature to 161°C, without aborting the

drying process.

Time interval for lowering the temperature to 160°C is 20 min.
For drying mode Quick overdriving Max drying temperature is 30% but no more

than Max default temperature.
For drying mode Step time interval for temperature maintenance is limited to 20

minutes for particular steps.


PRIOR VERSION OF MOISTURE ANALYZER
Parameter must be set to value <YES>.