Standard diaphragm replacement, 1 standard diaphragm replacement – Pulsafeeder Pulsar Shadow User Manual
Page 24

19
7.2.1
Standard Diaphragm Replacement
Figure 14
PULSAR diaphragms do not have a specific cycle life; however, the accumulation of foreign material
or the entrapment of sharp particles between the diaphragm and dish cavity can eventually cause
failure. Failure can also occur as a result of hydraulic system malfunction or chemical attack.
Periodic diaphragm inspection and replacement are recommended.
Diaphragm Replacement Procedure:
1. Disconnect the power source to the drive motor
2. Relieve all pressure from the piping system.
3. Take all precautions to prevent environmental and personnel exposure to hazardous materials.
4. Close the inlet and outlet shutoff valves.
5. Place a suitable container underneath the pump head to catch any liquid leakage.
6. Disconnect piping to the reagent head and drain any process liquid, following material safety
precautions described.
7. Remove all but one top reagent head bolt. Oil will leak out between the pump head and reagent head
as the bolts are loosened.
8. Tilt the head and pour out any liquids retained by the check valves into a suitable container,
continuing to follow safety precautions as appropriate.
9. Remove the final bolt and rinse or clean the reagent head with an appropriate material.
10. Remove and inspect the diaphragm. It may have taken a permanent convex/concave set as a result of
normal flexure and conformance to the dishplate. This condition is normal and is not cause for
replacement. The diaphragm must be replaced if it is deformed, dimpled, or obviously damaged.
11. To install a diaphragm, first ensure that the critical sealing areas of diaphragm, reagent head, and
pump head are clean and free of debris. Set the diaphragm in place on the reagent head and ensure
seating of the diaphragm sealing ring into the mating groove in the reagent head.
12. Install the reagent head bolts and tighten in an alternating pattern to ensure an even seating force.
Torque to the values recommended in Appendix IV.
13. Re-prime the pump head, see Section 7.2.2.
When reinstalling a used diaphragm it is not necessary to maintain the previous orientation relative
to the reagent head or pump head hole pattern.