Output Queue
Output Queue
The Output Queue stores response messages until they are read. If an unread 
message exists, Bit4 (Message Available 
− MAV) of the Status Byte Register is set
to 1. So, Bit4 is used to synchronize information exchange with the controller. See 
Figure 5-5.
Figure 5-5
Output Queue of B2200
The Response Formatter places Data Byte Messages and END messages into the 
Output Queue in response to query commands. These messages are removed from 
the Output Queue as they are read by the controller. As long as the Output Queue 
contains an unread message, MAV is 1.