Diagnostic:test:card:clear, Diagnostic:test:card[:execute – Agilent Technologies B2201A User Manual
Page 147

Agilent B2200 User’s Guide, Edition 2
5-21
SCPI Command Reference
:DIAGnostic:TEST:CARD:CLEar
:DIAGnostic:TEST:CARD:CLEar
This command clears the relay test result (pass/fail result) of the specified card.
Syntax
:DIAGnostic:TEST:CARD:CLEar card_number
Example
OUTPUT @Agb2200;":DIAG:TEST:CARD:CLE 1"
:DIAGnostic:TEST:CARD[:EXECute]?
This command executes the relay test for specified card, then returns the pass/fail
result. Before starting the relay test, open the input/output terminals (end of cable is
OK, Kelvin cable must be removed). For more information, see “RELAY_TEST”
on page 3-35.
NOTE
This command changes all settings of card to same state as after executing
:SYST:CPON command.
Syntax
:DIAGnostic:TEST:CARD[:EXECute]? card_number
Query response
test_result
1: failure card exists
0: pass
Example
OUTPUT @Agb2200;":DIAG:TEST:CARD? ALL"
Parameter Explanation
card_number
card number: 1, 2, 3, 4, or ALL
Parameter Explanation
card_number
card number: 1, 2, 3, 4, or ALL