System:cpon, System:cpon -46 – Agilent Technologies B2201A User Manual

Page 172

background image

5-46

Agilent B2200 User’s Guide, Edition 2

SCPI Command Reference
:SYSTem:CPON

:SYSTem:CPON

This command resets the specified card. Status after this command is shown below:

Channel Configuration:

Not changed.

Connection Rule:

Free

Connection Sequence:

Break Before Make

Bias Mode:

Off

Bias Input Port:

10

Bias-enabled Channels:

All channels (output ports) are bias-enabled.

Couple Mode:

Off

Couple Input Port:

Cleared.

Ground Mode:

Off

Ground Input Port:

12

Ground-enabled Channels:

Cleared.

Ground-enabled Ports:

Cleared.

Card Channel Status:

All relays are opened.

Self-Test Result:

Not changed.

Input Port Symbol String:

Not changed.

Output Port Symbol String:

Cleared.

For the device reset state, see *RST.

Syntax

:SYSTem:CPON card_number

Example

OUTPUT @Agb2200;":SYST:CPON ALL"

Parameter Explanation

card_number

For Auto Config mode: 0 or ALL
For Normal Config mode: 1, 2, 3, 4, or ALL