Channel configuration mode, Channel configuration mode -13 – Agilent Technologies B2201A User Manual
Page 73

Agilent B2200 User’s Guide, Edition 2
3-13
Front Panel Operation
Switch Control Functions
Channel Configuration Mode
The Agilent B2200 provides the two channel configuration modes, Normal and 
Auto. The configuration mode defines the way to control multiple switch modules 
installed in the mainframe.
Normal
Each module is considered independently. So each module is 
always a 12 output switching matrix.
The switch control functions need to be defined for each 
module.
Auto
Default mode at Power On or *RST command.
To use this mode, the same type of module must be installed in 
the mainframe from slot number 1 continuously (slots 1 to 2; or 
slots 1 to 3; or slots 1 to 4).
The installed modules are treated as one switching matrix. For 
example, if the modules are installed in slots 1 to 3, the Agilent 
B2200 works as a 36 output switching matrix.
The switch control functions are available for the switching 
matrix.