Installation
Output Connections
Table 2-4
Connection to the DUT Interface
CAUTION
Never connect the guard terminal to any output, including circuit common, chassis 
ground, or the guard terminal of any other unit. Doing so may result in an 
emergency condition.
For highly accurate current forcing and measurements while minimizing leakage, 
surround all force and sense lines from SMU by a guard as far as possible, and make 
cables stable by taping.
For reducing capacitance measurement error, tape the cables to any grounded 
materials such as shielding box. If you use probe card, grounding the probe card also 
reduces capacitance measurement error.
Kelvin connections
non-Kelvin connections
This connection is available only for the 
Kelvin connectors. 
This connection can cancel effects of cable 
resistance by connect sense line and force line 
as close as possible to DUT terminal.
Following connection is for the Kelvin 
connectors. For the triaxial connectors, 
ignore SENSE terminal, and make 
connection only for FORCE terminal. 
Measurement data will include residual 
resistance from the connection cable.