beautypg.com

Self-test, Self-test -10 – Agilent Technologies B2201A User Manual

Page 44

background image

2-10

Agilent B2200 User’s Guide, Edition 2

Installation
Self-Test

Self-Test

NOTE

To confirm the specifications

The self-test and diagnostics checks the operation of the mainframe and the
modules. However they cannot confirm if the Agilent B2200 satisfies its
specifications.

For verifying the specifications, contact your nearest Agilent Technologies Service
Center. Trained service personnel will perform calibration (performance
verification).

It is recommended to perform calibration once a year at least.

The following procedure performs the self-test and diagnostics.

1. Turn on the Agilent B2200.

The controller test will be performed. If the Agilent B2200 fails the test, contact
your nearest Agilent Technologies Service Center.

2. Press the

Shift

key and the

Menu

key in this order. Self-test menu will appear on

the LCD.

SELF_TEST

Controller test

RELAY_TEST

Relay test

KEY

Front panel interface test

BEEPER

Beeper test

LED

LED matrix test

PEN

Light pen test

GPIB

GPIB test

3. Move the cursor to the test item to perform, then press

Enter

key.

4. To start the test, move the cursor to EXECUTE, then press

Enter

key.

To display the previous test result, move the cursor to RESULT, then press

Enter

key.

For the details of each test, see “Selftest Menu” on page 3-34.