Agilent B2200 User’s Guide, Edition 2
7-7
Error Messages
Standard SCPI Error Messages
Execution Error
If syntax of a SCPI command header and parameter is valid, but the command 
cannot be executed due to some condition of the B2200, a -2XX error occurs. The 
error number and message are placed in the error queue, and bit4 of the Standard 
Event Status Register is set.
Error
Number
Error Message and Description
-200
Execution error
Generic execution error that cannot be determined more 
specifically.
-220
Parameter error
The parameter value is valid, but not executable due to some 
condition of the B2200. This error occurs if more specific errors 
-221 through -224 cannot be determined.
-222
Data out of range
A valid parameter setting could not be executed because interpreted 
value was out of range as defined by the B2200.
-223
Too much data
A valid parameter setting (block, expression, or string type) could 
not be executed because parameter contained more data than the 
B2200 could handle due to insufficient memory or other 
requirements.
-224
Illegal parameter value
A valid parameter setting (where exact value from list of possibles 
was expected) could not be executed due to present B2200 state.
-260
Expression error
An expression related error occurred.