Opc -8 – Agilent Technologies B2201A User Manual

Page 134

background image

5-8

Agilent B2200 User’s Guide, Edition 2

SCPI Command Reference
*OPC

*OPC

This command starts to monitor pending operations, and sets/clears the Operation
Complete (OPC) bit in the Standard Event Status Register as follows:

If there is no pending operation, sets the OPC bit to 1.

If there are any pending operations, sets the OPC bit to 0. The bit will be set to 1
again when all pending operations are completed.

So, *OPC command is required to enable the OPC bit. To stop monitoring pending
operations (disable OPC bit), execute the *CLS command.

For details about the Standard Event Status Register, see “Standard Event Status
Register” on page 5-58
. Also, see *WAI command.

Syntax

*OPC

Query response

1 <^END>

*OPC? places ASCII character 1 into the Output Queue when all pending
operations are completed. For details, see “Output Queue” on page 5-60.

Example

OUTPUT @Agb2200;"*OPC"
The following example is for query:

OUTPUT @Agb2200;"*OPC?"

ENTER @Agb2200;A