Agilent Technologies B2201A User Manual
Page 96

3-36
Agilent B2200 User’s Guide, Edition 2
Front Panel Operation
Setup Menus
KEY
Displays the sub menu that provides the following functions.
•
EXECUTE
Displays the following message.
DIAG:KEY
Press any key.
Press any front panel key and confirm the LCD display. For example, if you
press the
Port Function
key, the following message must be displayed.
DIAG:KEY
[Port Function]
Continue this for all keys. After that, press the
Enter
key twice for the normal
operation. The test result will be PASS.
If you find any wrong response, press the
Exit
key twice. The test result will be
FAIL.
Finally, press the
Exit
key to return to the previous menu.
•
RESULT
Displays the test result. Press the
Exit
key to return to the previous menu.
BEEPER
Displays the sub menu that provides the following functions.
•
EXECUTE
Starts the beeper test, and displays the following message.
DIAG:BEEP
Is beeper making 2 sounds?
Press the
Enter
key if the beeper works correctly. The test result will be recorded
as PASS.
If you find any problem, press the
Exit
key. The test result will be recorded as
FAIL.
Finally, press the
Exit
key to return to the previous menu.
•
RESULT
Displays the test result. Press the
Exit
key to return to the previous menu.