beautypg.com

Agilent Technologies B2201A User Manual

Page 222

background image

7-2

Agilent B2200 User’s Guide, Edition 2

Error Messages

This chapter lists and describes the error messages for Agilent B2200. An error
message consists of an error number and message.

There are two types of error messages:

Standard SCPI Error Messages

Negative error numbers (Command Error, Execution Error, Device-Dependent
Error, and Query Error) are standard SCPI errors.

B2200 Specific Error Messages

Positive error numbers are the B2200 specific errors.

Error messages are classified by error number as listed in the following table:

When an error occurs, the corresponding bit is set in the Standard Event Status
Register (see Chapter 5):

Also, the error number and message are placed in the error queue, which can be read
by the :SYSTem:ERRor? query command. The error queue is cleared by the
common command *CLS, and when power is turned on. For these commands, see
Chapter 5.

Error Range

Error Category

0

No error

-100 to -199

Command Error

-200 to -299

Execution Error

-300 to -399

Device-Dependent Error

-400 to -499

Query Error

100 to 32767

B2200 Specific Error

Error Category

Standard Event

Status Register Bit

Command Error

bit5

Execution Error

bit4

Device-Dependent Error

bit3

Query Error

bit2

B2200 Specific Error

bit3