beautypg.com

C.2.2 solvent flush, C.2.2 – Campbell Scientific EC155 CO2 and H2O Closed-Path Gas Analyzer and EC100 Electronics with Optional CSAT3A 3D Sonic Anemometer User Manual

Page 62

background image

Appendix C. EC155 Sample Cell and Intake Maintenance

3. Plug the hole in the inlet with your finger. The pump will pull a

vacuum on its internal filter/buffer volume, the pump tube, analyzer,

and intake tube.

4. After approximately one minute, unplug the hole.

During these steps, ambient air will rush in and blow dust from

the inner walls of the intake tube, which is likely be deposited on

the analyzer windows. For this reason, Campbell Scientific

recommends performing the dust blowout prior to cleaning

windows.

5. Repeat this procedure as needed.

When the windows no longer become dirty (CO

2

and H

2

O signal levels do not

change) this indicates no more dust is being removed from the intake tube.

C.2.2 Solvent Flush

If the intake tube has other contamination, such as salt deposits, it may be

flushed with water or alcohol, but be careful to keep the solvent out of the

pressure sensor passages (see earlier note on cleaning the sample cell).

Following the steps in the following procedure, will help keep the pressure

sensor passages clear.

1. Power the system down.
2. Remove the sensor head from the sample cell.
3. Close the lid of the sample cell assembly.
4. Position sample cell assembly upside down. This allows the solvent

to flow away from the pressure sensor passage.

5. Tilt slightly so that the intake is higher than the sample cell.
6. Remove the intake filter.
7. Fill a syringe with solvent (generally, tap water will be sufficient) and

press it against the hole in the end of the intake tube.

8. Depress the plunger to let the solvent flow through the intake tube.

The waste solvent will collect in the lid of the sample cell assembly.

9. Fill the syringe with air and push the air through the intake tube to

force most of the solvent out of the tube.

10. Dump the solvent out of the sample cell assembly and wipe the

assembly dry.

11. Clean the analyzer windows and the sample cell as described in

Section 9.3, Cleaning Analyzer Windows, and C.1, Cleaning Sample

Cell, in the section above.

This procedure is likely to leave some of the solvent in the system.

Make sure it is completely dry before attempting a zero/span.

NOTE

NOTE

C-2