Appendix A. Filter Bandwidth and Time Delay

FIGURE A-2. Frequency response comparison of the EC100 10-Hz

bandwidth and a 50-msec moving average

TABLE A-1. Filter Time Delays for Various Bandwidths

Bandwidth (Hz)

Time Delay (ms)

5

800

10

400

12.5

320

20

200

25

160