beautypg.com

3 example crbasic program, Example crbasic program – Campbell Scientific EC155 CO2 and H2O Closed-Path Gas Analyzer and EC100 Electronics with Optional CSAT3A 3D Sonic Anemometer User Manual

Page 53

background image

EC155 CO

2

and H

2

O Closed-Path Gas Analyzer

10.3 Example CRBasic Program

'CR3000 Series Datalogger

'CR3000 Series Datalogger

Public sonic_irga(13)

Alias sonic_irga(1) = Ux

Alias sonic_irga(2) = Uy

Alias sonic_irga(3) = Uz

Alias sonic_irga(4) = Ts

Alias sonic_irga(5) = diag_sonic

Alias sonic_irga(6) = CO2

Alias sonic_irga(7) = H2O

Alias sonic_irga(8) = diag_irga

Alias sonic_irga(9) = cell_tmpr

Alias sonic_irga(10) = cell_press

Alias sonic_irga(11) = CO2_sig_strgth

Alias sonic_irga(12) = H2O_sig_strgth

Alias sonic_irga(13) = diff_press

Units Ux = m/s

Units Uy = m/s

Units Uz = m/s

Units Ts = C

Units diag_sonic = arb

Units CO2 = umol/mol

Units H2O = mmol/mol

Units diag_irga = arb

Units cell_tmpr = C

Units cell_press = kPa

Units CO2_sig_strgth = arb

Units H2O_sig_strgth = arb

Units diff_press = kPa

DataTable (ts_data,TRUE,-1)

DataInterval (0,0,mSec,10)

Sample (13,Ux,IEEE4)

EndTable

BeginProg

Scan (100,mSec,0,0)

EC100 (Ux,1,2)

CallTable ts_data

NextScan

EndProg

43