1. output modes for ec100 instruction, 2. bits in the sonic diagnostic flag, Ee table 10-2 – Campbell Scientific EC155 CO2 and H2O Closed-Path Gas Analyzer and EC100 Electronics with Optional CSAT3A 3D Sonic Anemometer User Manual

Page 48

background image

EC155 CO

2

and H

2

O Closed-Path Gas Analyzer

TABLE 10-1. Output Modes for EC100 Instruction

Output

Mode

Data

Field

Description

Units

0, 1, 2,

1

Ux

m/s

2

Uy

m/s

3

Uz

m/s

4

Sonic Temperature

ºC

5

Sonic Diagnostic Flag

6

CO

2

µmol/mol

7

H

2

O

mmol/mol

8

Gas Diagnostic Flag

1, 2

9

Air Temperature

ºC

10

Air Pressure

kPa

11

CO

2

Signal Strength

nominally 0.0 ≤ strength

≤1.0

12

H

2

O Signal Strength

nominally 0.0 ≤ strength

≤1.0

2

13

Sample Cell Pressure

Differential

kPa

As shown, all output modes give two diagnostic values, the

Sonic Diagnostic

Flag and the Gas Diagnostic Flag. The values contain a bit field, with each

bit representing a monitored condition. When a certain condition is detected,

the corresponding bit is set. The value remains set until the event that caused

the condition is no longer present. TABLE 10-2 and TABLE 10-3 below

describe the bits in the sonic diagnostic flag and the gas diagnostic flag,

respectively.

TABLE 10-2. Bits in the Sonic Diagnostic Flag

bit

hex

value

decimal

Name

Function

0

0x1

1

Low Amp

Amplitude is too low

1

0x2

2

High Amp

Amplitude is too high

2

0x4

4

Tracking

Poor signal lock

3

0x8

8

Hi 3 Axis DC

Delta temperature exceeds

limits

4

0x10

16

Acquiring

Acquiring ultrasonic

signals

5

0x20

32

Cal Mem Err

Sonic head calibration

signature error

38