EC155 CO

2

and H

2

O Closed-Path Gas Analyzer

FIGURE 6-3. Exploded view of mounting the EC100 enclosure

h. Remove the EC100 enclosure desiccant from the plastic bag and put it

back in the mesh pocket of the enclosure. Adhere the humidity indicator

card to the inside of the enclosure.

6.2 Plumbing

6.2.1 Flow

The EC155 has a small sample cell volume (5.9 cm

3

) to give good frequency

response at a relatively low flow rate. The sample cell residence time is 50 ms

for a nominal 7 LPM flow. The CPEC200 pump module is designed to provide

this flow for the EC155, but other user-supplied pumps may be used. There is

no specific limitation to the flow rate that may be used with the EC155, but the

sample cell pressure must be considered.

6.2.2 Pressure

The EC155 is designed to be used near ambient pressure, but it will not be

damaged by operation under vacuum. The EC155 includes a differential

pressure sensor to measure the sample cell pressure relative to ambient

pressure, which has a range of ± 7 kPa. If the EC155 is operated less than

7 kPa from ambient pressure, the user must attach a separate, user-supplied

pressure sensor.

11