beautypg.com

Appendix a. filter bandwidth and time delay – Campbell Scientific EC155 CO2 and H2O Closed-Path Gas Analyzer and EC100 Electronics with Optional CSAT3A 3D Sonic Anemometer User Manual

Page 55

background image

Appendix A. Filter Bandwidth and Time

Delay

The EC100 measures CO

2

and H

2

O from the EC155 gas analyzer head (as well

as wind velocity and sonic temperature from the optional CSAT3A sonic head)

at 100 Hz and then applies a user-selectable low-pass filter. The available filter

bandwidths are 5, 10, 12.5, 20, and 25 Hz. FIGURE A-1 shows the amplitude

response of these filters. The EC100 filters provide a flat pass band, a steep

transition from pass band to stop band, and a well-attenuated stop band.

FIGURE A-2 compares the EC100 10-Hz filter to a 50-msec moving average

filter with approximately the same bandwidth.

The ideal eddy-covariance filter is one that is wide enough to preserve the low-

frequency signal variations that transport flux and narrow enough to attenuate

high-frequency noise. In addition, to minimize aliasing (the misinterpretation

of high-frequency variation as lower-frequency variation) the measurement

bandwidth must be less than half of the sample rate, or the datalogger scan rate.

Two factors complicate choosing the ideal eddy-covariance bandwidth. First,

the flux signal bandwidth varies from one installation to another, and the flux

signal bandwidth varies with mean wind speed at a given installation. Second,

the fast sample rate required to anti-alias a desired signal bandwidth may result

in large, unwieldy data sets.

Fortunately, the covariance calculation itself relaxes the need for the ideal

bandwidth. First, the time-averaged (typically thirty-minute) covariance

calculations inherently reduce noise, and second, aliasing does not degrade the

accuracy of covariance calculations. Therefore, the factory default for the

EC100 bandwidth (20 Hz) is rather wide to preserve the signal variations that

transport flux, and that bandwidth is suitable for most flux applications.

Additional bandwidths are available for experimenters desiring to match the

EC100 filter bandwidth to their data acquisition sample rate to avoid aliasing.

In this case, the selected bandwidth should be one-half of the sample rate (or

datalogger scan rate), and experimenters should be careful to avoid attenuation

of flux-carrying signals.

The EC100 electronics synchronously sample the gas in the EC155 sample cell

and the CSAT3A sonic head. However, delays induced by the intake assembly

must be accounted for. The exact delay will depend on the length and size of

the intake tubing and the pump flow rate. This delay needs to be

experimentally determined by shifting the time delay until the covariance of

the vertical wind and the gas concentrations are maximized.

Experimenters wishing to synchronize their EC100 data with other

measurements in the data acquisition system must account for the time delay of

the EC100 filter. TABLE A-1 shows the delay for each of the filter

bandwidths. The EC100 provides a constant time delay for all spectral

components within each filter’s pass band.

The following examples show how to use TABLE A-1. To synchronize

EC100 data to other datalogger measurements when the datalogger scan rate is

25 Hz and the EC100 bandwidth is set to 20 Hz (a 200-msec time delay from

TABLE A-1), delay the non-EC100 data by five datalogger scans. Similarly,

for a 10-Hz datalogger scan rate and the same 20-Hz EC100 bandwidth, delay

A-1