beautypg.com

Zero span span gain – Campbell Scientific EC155 CO2 and H2O Closed-Path Gas Analyzer and EC100 Electronics with Optional CSAT3A 3D Sonic Anemometer User Manual

Page 43

background image

EC155 CO

2

and H

2

O Closed-Path Gas Analyzer

i. Examine the measurements that were written down for span CO

2

, span

H

2

O, and zero air. Compute the drift in instrument gain using the

following equation:

meas

meas

actual

zero

span

span

gain

=

where,

span

actual

= known concentration of the span gas

span

meas

= measured concentration of the span gas

zero

meas

= measured concentration in zero gas.

Note that in the zero-and- span window of ECMon, span

actual

is reported to

the right of the box where the user enters the span dew-point temperature.

The software calculates span

actual

by taking into account the dew-point

temperature and current ambient temperature and pressure. The equations

used for this calculation may be found in Appendix B. If drift (offset or

gain) for CO

2

or H

2

O is excessive, it may be time to replace the desiccant

and CO

2

scrubber bottles (see Section 9.5, Replacing Desiccant and

Scrubber Bottles).

j. With the zero air still flowing and measurements stabilized, click on the

Zero CO

2

and H

2

O button in the Zero/Span window. This will cause the

analyzer to adjust the values of its CO

2

Zero and H

2

O Zero parameters,

forcing the CO

2

and H

2

O concentrations to read zero. Verify the CO

2

and

H

2

O concentrations now read zero.

k. Now, remove the zero air source and replace it with the CO

2

span gas.

Allow the gas to flow through the sample cell. Watch for readings to

stabilize.

33