beautypg.com

3 cleaning analyzer windows, Cleaning analyzer windows, 1. the underside of the optional heated intake – Campbell Scientific EC155 CO2 and H2O Closed-Path Gas Analyzer and EC100 Electronics with Optional CSAT3A 3D Sonic Anemometer User Manual

Page 38: Figure 9-1)

background image

EC155 CO

2

and H

2

O Closed-Path Gas Analyzer

FIGURE 9-1. The underside of the optional heated intake

9.3 Cleaning Analyzer Windows

The windows of the analyzer should be cleaned if the signal strength of CO

2

or

H

2

O drops below 80% of the original value. (These values may be monitored

in the output data, or they can be viewed with ECMon.) To clean the windows,

follow these steps:

a. Stop the air flow through the EC155.

b. Loosen the two captive thumbscrews (one on each end of the EC155), and

lift the top portion of the EC155 shell, leaning it back against the lower

shell. See FIGURE 9-2.

c. Loosen the thumbscrew on the cable clamp at the back of the analyzer to

release the cable, and loosen the two long thumbscrews found above the

sample cell. Rotate the latches so that the struts on the analyzer are free to

move upwards. See FIGURE 9-3.

d. Lift the analyzer head off the sample cell (see FIGURE 9-4), taking care

not to lose the O-rings (pn 26212) surrounding the optical windows (see

FIGURE 9-5). If an O-ring is lost, two replacement O-rings may be found

in the mesh pocket of the EC100 enclosure, or new ones may be ordered.

e. Wash the windows with isopropyl alcohol using cotton swabs or a non-

scratching tissue or cloth.

f. Put the analyzer back in place, making sure the O-rings are still intact.

The analyzer’s label should face out to the side.

g. Rotate the latches back in place to hold the analyzer’s struts down, and

tighten the long thumbscrews by hand. Also make sure the analyzer cable

is seated properly in the cable clamp and tighten the thumbscrew by hand.

h. Put the top portion of the EC155 shell back in place, and tighten the

thumbscrews.

Santoprene Tab

28