Replacing the ec155 desiccant/co, Scrubber bottles, 6. ecmon zero/span window – Campbell Scientific EC155 CO2 and H2O Closed-Path Gas Analyzer and EC100 Electronics with Optional CSAT3A 3D Sonic Anemometer User Manual

Page 44: N 9.5, 5 replacing the ec155 desiccant/co

background image

EC155 CO

2

and H

2

O Closed-Path Gas Analyzer

FIGURE 9-6. ECMon Zero/Span window

l. Enter the known concentration of CO

2

(in ppm) in the

Span

Concentration box and press Span. This will cause the analyzer to adjust

the value of its CO

2

Span parameter, forcing the measured CO

2

concentration to the value specified. Verify the CO

2

concentration reads

the correct value.

m. Replace the CO

2

span gas with an H

2

O span gas of known dew point.

Allow the gas to flow through the sample cell; as before, higher flows may

be desired for a couple minutes to more quickly establish equilibrium

before resuming a flow between 0.2 and 0.4 L/min. Wait for the readings

to stabilize.

n. Enter the known dew point (in °C) in the

Span Dew Point box and press

Span. This will cause the analyzer to adjust the value of its H

2

O Span

parameter, forcing the measured dew point to the value specified. Verify

the dew point reads the correct value.

o. The zero-and-span procedure is now complete.

9.5 Replacing the EC155 Desiccant/CO

2

Scrubber Bottles

If more than one year has passed since replacing the desiccant/scrubber, or if

zero-and-span readings have drifted excessively (see Section 9.4, Zero and

Span above), the desiccant/scrubber bottles (pn 26511) within the EC155

analyzer head should be replaced as follows:

a. Remove the analyzer in the same way as explained in Section 9.3,

Cleaning Analyzer Windows of this manual.