7.2.5 RS-485 Baud Rate

If the unprompted output mode is set to RS-485, this parameter determines the

baud rate. Otherwise this setting is not used. The RS-485 baud rate defaults to

115200 bps, although the user may enter another value.

7.2.6 Analog Output

The EC100 has two analog outputs for CO

2

and H

2

O molar mixing ratios (see

Section 8.3, Analog Outputs for more information). These outputs may be

enabled/disabled with this setting. The default is for analog output to be

disabled.

7.2.7 ECMon Update Rate

This setting determines the rate at which data are sent over the USB connection

to the PC while running ECMon. The default setting of 10 Hz should be

adequate in most situations.

7.2.8 Temperature Sensor

The EC155 measures the temperature of the sample cell block with a

thermocouple embedded in the block. With the

Auto-Select default setting,

the EC100 will automatically detect that an EC155 is connected to the

electronics and will report temperature measurements from the sample cell

thermocouple.

To diagnose problems with the temperature measurement, a fixed temperature

value may be used, or the temperature sensor may be selected manually.

7.2.9 Fixed Temperature Value

If the

Temperature Sensor setting is None, the EC155 will use the value of

this setting for the sample-cell temperature. This mode is intended for

troubleshooting only. In normal operation, the

Temperature Sensor is set to

Auto-Select, and this setting is not used.

7.2.10 Pressure Sensor

This setting determines which pressure sensor will be used to measure the

barometric pressure. The EC100 always includes the EC100 basic barometer,

but it may be ordered with the optional EC100 enhanced barometer. This

setting is factory defaulted to the enhanced barometer if it is ordered, and to the

basic barometer otherwise.

There are two other possible settings for the

Pressure Sensor. First, the user

may provide his or her own pressure sensor. In this case the setting should be

changed to

User Supplied, with the appropriate values for gain and offset

entered (see below). This option may be used if the EC155 sample cell is to be

used outside the range of the differential pressure sensor (see Section 7.2.11)

For this mode the user-supplied pressure sensor must be plumbed to the EC155

sample cell, and the

Differential Pressure sensor setting should be disabled.

19