beautypg.com

4 zero and span, Zero and span – Campbell Scientific EC155 CO2 and H2O Closed-Path Gas Analyzer and EC100 Electronics with Optional CSAT3A 3D Sonic Anemometer User Manual

Page 41

background image

EC155 CO

2

and H

2

O Closed-Path Gas Analyzer

9.4 Zero and Span

As is the case with optical instrumentation, the EC155 may drift slightly with

exposure to natural elements. Thus, a zero-and-span procedure should be

performed occasionally. The first part of the procedure listed below simply

measures the CO

2

and H

2

O span and zero, without making any adjustments.

This allows the CO

2

and H

2

O gain factors to be calculated. These gain factors

quantify the state of the analyzer before the zero-and-span procedure, and in

theory could be used to correct recent measurements for drift. The last part of

the zero-and-span procedure adjusts internal processing parameters to correct

subsequent measurements.

If the EC155 was purchased as part of a CPEC200 closed-path

eddy covariance system, consult the CPEC200 manual. The

CPEC200 system has an optional valve module to allow the

datalogger to automate the zero-and-span procedure.

It is imperative that the zero-and-span procedure be done correctly and not

rushed; allocate plenty of time for the procedure. During a normal zero-and-

span procedure a PC running the ECMon software is used to monitor and

control the EC155. However, the zero-and-span procedure can also be

performed using either the Device Configuration Utility software or a

datalogger running the

EC100Configure() instruction (see Section 10.2,

EC100Configure() Instruction).

To check and then set the EC155 zero and span, follow the steps below:

a. Connect the EC100 to a PC with the EC100 USB cable (pn 26563), and

launch ECMon on the PC. Select the appropriate USB port, and press

Connect. The main screen should now be reporting real-time CO

2

and

H

2

O concentrations.

b. Check the differential pressure and replace the intake filter as needed (see

Section 9.2, Intake Filter Replacement).

c. Check the signal strengths and clean the windows as needed (see Section

9.3, Cleaning Analyzer Windows).

d. Verify that the Gas LED status light on the EC100 panel is green. Also

make sure the analyzer is resting right-side-up. If the zero-and-span

procedure is being performed on-site, this should already be the case.

e. If the EC155 is configured with the heated intake assembly, connect the

zero-and-span gas to the Zero/Span inlet at the back of the analyzer. The

zero-and-span gas will be pushed backwards through the EC155 sample

cell and exhausted through the heated intake assembly. If the sample

pump is the CPEC200 pump module, it may be left connected and simply

shut off. A fraction of the zero-and-span gas will be pushed through the

pump, but not enough to affect the equilibration time. Alternatively, the

sample pump may be disconnected and the Pump connection plugged.

If the EC155 is configured with a sample inlet fitting to connect to the

user’s own intake assembly, there are two options for connecting the zero-

and-span gas:

NOTE

31