2 output signals, Output signals – Campbell Scientific EC155 CO2 and H2O Closed-Path Gas Analyzer and EC100 Electronics with Optional CSAT3A 3D Sonic Anemometer User Manual

Page 15

background image

EC155 CO

2

and H

2

O Closed-Path Gas Analyzer

1

noise rms, assumes:

o 25°C

o 85 kPa

o 19 mmol/mol H

2

O concentration

o 326 mmol/mol CO

2

concentration

o 25 Hz bandwidth.

2

user selectable

3

-30° to 50°C

4

noise rms

5

assumes:

o -30° to +50°C

o wind speed <30 m·s

-1

o azimuth angles between ±170°

6

refer to manufacturer’s product brochure or manual for details

5.2 Output Signals

Features

EC100 electronics outputs data using:

o CS SDM

o RS-485

o USB

o Analog out

Digital

SDM (Synchronous Device for Measurement)

1

Data type:

FLOAT

RS-485

Data type:

ASCII

Output Rate

2

:

5 to 50 Hz

Baud rate

2

:

1200 to 230400 bps

USB

Data type:

ASCII

Output rate

2

:

10, 25 or 50 Hz

Analog (two outputs for CO

2

and H

2

O molar mixing ratios)

Voltage range:

0 mV to 5000 mV

Resolution:

76 µV (16 bit)

Update rate:

150 Hz

Accuracy (at 25

o

C):

±3 mV

CO

2

mixing ratio equation: µmol/mol = 211.27 (V

out

) – 56.34

Full scale range:

-56 to 1000 µmol/mol

H

2

O mixing ratio equation: mmol/mol = 11.31 (V

out

) – 3.04

Full scale range:

-3 to 53 mmol/mol

1

Synchronous Device for Measurement. A Campbell Scientific, Inc. proprietary serial interface

for datalogger to peripheral and sensor communication. See Section 8.1, SDM Output for details.