beautypg.com

Campbell Scientific EC155 CO2 and H2O Closed-Path Gas Analyzer and EC100 Electronics with Optional CSAT3A 3D Sonic Anemometer User Manual

Page 42

background image

EC155 CO

2

and H

2

O Closed-Path Gas Analyzer

1. Connect the zero-and-span gas to the Sample inlet, and disconnect the

pump, leaving the Pump connection open. The zero-and-span gas will

be pushed forward through the EC155 sample cell and exhausted out

the Pump fitting. In this case the Zero/Span connection may be left

plugged.

2. Connect the zero-and-span gas to the Zero/Span inlet, and disconnect

the intake tube from the Sample connection. Disconnect the sample

pump and plug the Pump connection. The zero-and-span gas will be

pushed backwards through the EC155 sample cell and exhausted

through the Sample fitting.

f. Click on the

Zero/Span button on the main screen of ECMon. A real-time

graph at the bottom of the window will appear that displays concentrations

of CO

2

and H

2

O (see FIGURE 9-6). Next allow CO

2

span gas to flow

through the sample cell. The exact flow rate is not important since sample

cell pressure is being measured; however, a flow rate should be high

enough to flush the tubing and sample cell in a reasonable time. If the

tubing from the CO

2

span tank to the EC155 is kept short, the CO

2

will

equilibrate in several seconds even at relatively low flow rate (< 0.5 LPM).

However if the tubing is long (e.g. if the EC155 is left in place at the top of

the tower) it may take a few minutes to flush the tube, and a higher flow

rate (> 1 LPM) may be useful to reduce the equilibration time.

Once gas begins to flow through the sample cell, watch the graph on

ECMon for the measurement readings to stabilize. Once stable, write

down the reported CO

2

concentration.

Optimally the concentration of span CO

2

should be near the

concentration of CO

2

being measured in the field. Also, the user

is advised to use CO

2

mixtures in dry ambient air for the CO

2

span

gas. The use of reference CO

2

gas mixtures in pure nitrogen will

lead to errors due to a carrier gas effect on pressure-broadening of

the CO

2

absorption lines since oxygen gas has a smaller line-

broadening coefficient than nitrogen.

g. Stop the flow of CO

2

span gas, and provide H

2

O span gas to the analyzer.

A dew point generator is often used for this. Allow a relatively high flow

rate for the first several minutes to more quickly stabilize the system, and

then decrease the flow to 0.2 to 0.4 L/min before making the measurement.

Higher flow rates should not be used when taking the measurement

because back-pressure on the dew point generator will cause errors. Write

down the reported H

2

O concentration.

As H

2

O may adsorb to surfaces inside the tubing and sample cell,

allow plenty of time for the system to reach equilibrium.

h. Stop the flow of H

2

O span gas, and allow zero air (no CO

2

or H

2

O) to flow

through the analyzer. Dry nitrogen is often used as zero air. The exact

flow rate is not important since sample cell pressure is being measured,

however, a flow rate should be high enough to flush the tubing and sample

cell within a reasonable time period. Wait for the measurement readings

to stabilize and write down the reported values for CO

2

and H

2

O

concentrations.

NOTE

NOTE

32