EC155 CO

2

and H

2

O Closed-Path Gas Analyzer

Factory calibrated range

CO

2

:

0 to 1000 µmol·mol

-1

H

2

O:

0 mmol·mol

-1

to 37

o

C dewpoint

Analyzer temp:

-30

o

to 50

o

C

Baro pressure:

70 to 106 kPa

CO

2

performance

Zero max drift

3

:

±0.55 mg·m

-3

·°C

-1

(±0.3 μmol·mol·°C

-1

)

Gain Drift:

±0.1% of reading·°C

-1

(maximum)

Sensitivity to H

2

O:

±5.6 x 10

-5

µmol CO

2

·mol

-1

H

2

O (max)

H

2

O performance

Zero max drift

3

:

±0.037 g·m

-3

·°C

-1

(±0.05 mmol·mol

-1

·°C

-1

)

Gain Drift:

±0.3% of reading·°C

-1

(maximum)

Sensitivity to CO

2

:

±0.05 mol H

2

O·mol

-1

CO

2

(maximum)

CSAT3A sonic measurement precision

4

u

x

:

1 mm·s

-1

u

y

:

1 mm·s

-1

u

z

:

0.5 mm·s

-1

Sonic temperature:

0.025

o

C

CSAT3A sonic accuracy

5

Offset error

u

x

, u

y

:

< ±8 cm·s

-1

u

z

:

< 4 cm·s

-1

Gain error

Wind vector ±5° horizontal: < ±2% of reading

Wind vector ±10° horizontal: < ±3% of reading

Wind vector ±20° horizontal: < ±6% of reading

CSAT3A sonic reporting range

Full scale wind:

±65.553 m/s

Sonic temperature:

-50° to +60°C

Sample cell sensors

6

Barometer

Basic barometer

Accuracy

-30 to 0°C:

±3.7 kPa at -30°C, falling linearly to ±1.5

kPa at 0

o

C

0° to 50°C:

±1.5 kPa

Measurement rate: 10 Hz

Optional enhanced barometer:

Manufacturer:

Vaisala

Model:

PTB110

Accuracy

-30 to 0°C:

±0.15 kPa

Measurement rate: 1 Hz

Temperature sensor:

Manufacturer: BetaTherm

Model:

100K6A1A Thermistor

Accuracy:

±0.15

o

C (-30

o

to 50

o

C)

4