beautypg.com

1 pressure gain, 2 pressure offset, 3 fixed pressure value – Campbell Scientific EC155 CO2 and H2O Closed-Path Gas Analyzer and EC100 Electronics with Optional CSAT3A 3D Sonic Anemometer User Manual

Page 30: 11 differential pressure, 3 ecmon, Pressure gain, Pressure offset, Fixed pressure value, Differential pressure, Ecmon

background image

EC155 CO

2

and H

2

O Closed-Path Gas Analyzer

The final option is to select

None for the Pressure Sensor setting. The EC100

will use a fixed (see below) value for pressure. This mode is intended for

troubleshooting only.

7.2.10.1 Pressure Gain

If the

Pressure Sensor is set to User Supplied, this setting gives the gain

factor (kPa/V) used to convert measured voltage to pressure. Normally the

Pressure Sensor is set to EC100 Basic or EC100 Enhanced, and this setting

is not used.

7.2.10.2 Pressure Offset

If the

Pressure Sensor is set to User Supplied, this setting gives the offset

(kPa) used to convert measured voltage to pressure. Normally the

Pressure

Sensor is set to EC100 Basic or EC100 Enhanced, and this setting is not

used.

7.2.10.3 Fixed Pressure Value

If the

Pressure Sensor setting is None, the EC155 will use the value of this

setting for the barometric pressure. This mode is intended for troubleshooting

only. In normal operation this setting is not used.

7.2.11 Differential Pressure

The EC155 includes a differential pressure sensor to measure the pressure

difference between the inside of the sample cell and barometric pressure. With

the

Auto-Select default, the EC100 detects the presence of the EC155 and

automatically enables the differential pressure measurement. This pressure

difference is then added to the barometric pressure measurement to give the

pressure in the sample cell.

The EC155 sample cell differential pressure sensor has a range of ± 7 kPa,

which will accommodate most applications. If the sample cell is to be used

outside this range, the user must disable the

Differential Pressure sensor and

connect a user-supplied pressure sensor (see Section 7.2.10, Pressure Sensor).

7.3 ECMon

Settings for the EC155 are easily verified and/or changed by using the

Windows PC support software ECMon (ECMon is short for Eddy Covariance

Monitor), which is found on the EC150 & EC155 Support CD (pn 27007) or

on the company website in the Support|Downloads section

(

www.campbellsci.com/downloads

).

Before installing ECMon, read the file titled Read.me.text found on the EC150

& EC155 Support CD. This will direct the user to install USB drivers (also

found on the Support CD), which are required for communications between the

PC and the EC100 via the EC100 USB cable (pn 26563). Once the drivers are

installed, download and run the ECMon.exe install file. Launch ECMon, and

connect the EC100 electronics to the PC with the included EC100 USB cable

(pn 26563). The USB connection for the EC100 electronics is found on the

bottom of the enclosure (see FIGURE 6-6). Once connected, select the

appropriate communications port in the ECMon Main Page and click

Connect

(see FIGURE 7-1). Next click on the

Setup button. All of the above settings

are now available for the user to change (see FIGURE 7-2).

20