Appendix 1: Serial commands
93
The processor does not allow the following combinations:
•
no parity, 7 data bits, 1 stop bit: if this combination is given the
HMP230 programme will change the number of stop bits to 2
•
even or odd parity, 8 data bits, 2 stop bits: if this combination is given
the programme changes the number of stop bits to 1
NOTE
The serial bus settings become effective only after reset.
When the half-duplex mode is set, it will automatically turn the echo off. Even
then the ECHO command can indicate that echo is on.
UNIT
Selecting the output units
UNIT x
x =
m(etric units)
n(on-metric units)
quantity
metric
non-metric
RH = relative humidity
%RH
%RH
T = temperature
°C
°F
Td = dewpoint temperature
°C
°F
a = absolute humidity
g/m
3
gr/ft
3
x = mixing ratio
g/kg
gr/lb
h = enthalpy
kJ/kg
Btu/lb
Tw = wet bulb temperature
°C
°F
For example, the command for setting the non-metric units is:
>UNIT N
Output units : non metric
When the command is given with no parameters, the transmitter outputs the
currently valid setting.