M210225en-B

Appendix 1: Serial commands

93

The processor does not allow the following combinations:

no parity, 7 data bits, 1 stop bit: if this combination is given the
HMP230 programme will change the number of stop bits to 2

even or odd parity, 8 data bits, 2 stop bits: if this combination is given
the programme changes the number of stop bits to 1

NOTE

The serial bus settings become effective only after reset.

When the half-duplex mode is set, it will automatically turn the echo off. Even
then the ECHO command can indicate that echo is on.

UNIT

Selecting the output units

UNIT x

x =

m(etric units)
n(on-metric units)

quantity

metric

non-metric

RH = relative humidity

%RH

%RH

T = temperature

°C

°F

Td = dewpoint temperature

°C

°F

a = absolute humidity

g/m

3

gr/ft

3

x = mixing ratio

g/kg

gr/lb

h = enthalpy

kJ/kg

Btu/lb

Tw = wet bulb temperature

°C

°F

For example, the command for setting the non-metric units is:

>UNIT N
Output units : non metric

When the command is given with no parameters, the transmitter outputs the
currently valid setting.