HMP230 SERIES
Appendix 1: Serial commands

M210225en-B

82

ASEL

Scaling the analogue outputs

Disconnect the security lock jumper!

ASEL xxx yyy

xxx

=

channel 1's quantity

yyy

=

channel 2's quantity:
RH

= relative humidity

T =

temperature

Td

= dewpoint temperature

Abs = absolute humidity
Mix = mixing ratio
h =

enthalpy

Tw

= wet bulb temperature

Example: relative humidity selected on channel 1 and temperature on channel
2

>ASEL RH T
Ch1

(RH) lo

0.000 %RH?

Ch1

(RH) hi

100.000 %RH

?

Ch2

(T

)

lo

-40.000 'C

?

Ch2

(T

)

hi

+160.000 'C

?

When the ASEL command is given on its own, the transmitter outputs its cur-
rent settings:

>ASEL
Ch1

(RH) lo

0.000 %RH

Ch1

(RH) hi

100.000 %RH

Ch2

(T ) lo

-40.000 'C

Ch2

(T ) hi

+160.000 'C

The outputs and their scales can also be given directly with the ASEL com-
mand.

ASEL xxx yyy aaa.a bbb.b ccc.c ddd.d

xxx

=

channel 1's quantity

yyy

=

channel 2's quantity

aaa.a

=

lower limit of channel 1

bbb.b

=

upper limit of channel 1

ccc.c

=

lower limit of channel 2

ddd.d

=

upper limit of channel 2