Vaisala HMP230 User Manual
Page 43

HMP230 SERIES
M210225en-B
User's Guide
37
6.3.4. Analogue output commands
6.3.4.1. Selecting the output (mA/V)
Disconnect the security lock jumper!
•
Select Mode in the main menu and Analog outputs in the Mode
menu:
•
Select Mode ( mA / V ). The current settings for channel 1 are dis-
played:
•
If the settings are correct, press ENT.
•
If the settings need to be changed, press CL:
−
the quantity (mA/V) starts flashing; it can be changed with the ar-
row keys and acknowledged with the ENT key
−
the lower limit starts flashing
−
acknowledge the lower limit with ENT or start changing it by
pressing CL; a new lower limit is given one character at a time
with the arrow keys
−
the upper limit starts flashing
•
acknowledge the upper limit with ENT or start changing it by pressing
CL; a new upper limit is given one character at a time with the arrow
keys
When channel 1 has been set, the program goes on to channel 2; the procedure
is the same as with channel 1.
NOTE
Also the analogue output jumpers must be set to correct
places (see Figure 23).