4800 baud, Even parity, 7 data bits – Vaisala HMP230 User Manual
Page 34: 1 stop bit, Full duplex
HMP230 SERIES
User's Guide
M210225en-B
28
2
5
3
3
7
2
3
7
2
TX
GND
RX
TX
GND
RX
TX
GND
RX
HMP 230
D9S
D25S
D25P
PC
TXD
RXD
TXD
TXD
RXD
RXD
4
6
7
8
5
6
8
20
TERMINAL
Figure 26
Connection of cables
When the serial bus has been connected between the PC and the transmitter,
the PC is switched on. When using a PC, a terminal emulation program (e.g.
Procomm Plus, Datastorm or Windows terminal) is started.
The factory settings for data transfer are:
•
4800 baud
•
even parity
•
7 data bits
•
1 stop bit
•
full duplex
NOTE
When the serial bus settings are changed, the transmit-
ter has to be reset before the new settings become ef-
fective.
The processor does not allow the following combinations:
•
no parity, 7 data bits, 1 stop bit: if this combination is given the
HMP230 program will change the number of stop bits to 2
•
even or odd parity, 8 data bits, 2 stop bits: if this combination is given
the program changes the number of stop bits to 1
Refer to the manuals of the PC and the terminal emulation program when
giving serial settings.
The RS 232C screw terminal cannot be used if an RS 485/422 serial module
or a current loop module is used. See Appendices 3 and 4 on how to install
and operate these modules.