beautypg.com

4800 baud, Even parity, 7 data bits – Vaisala HMP230 User Manual

Page 34: 1 stop bit, Full duplex

background image

HMP230 SERIES
User's Guide

M210225en-B

28

2
5
3

3
7
2

3
7
2

TX
GND
RX

TX
GND
RX

TX
GND
RX

HMP 230

D9S

D25S

D25P

PC

TXD

RXD

TXD

TXD

RXD

RXD

4
6
7
8

5
6
8

20

TERMINAL

Figure 26

Connection of cables

When the serial bus has been connected between the PC and the transmitter,
the PC is switched on. When using a PC, a terminal emulation program (e.g.
Procomm Plus, Datastorm or Windows terminal) is started.

The factory settings for data transfer are:

4800 baud

even parity

7 data bits

1 stop bit

full duplex

NOTE

When the serial bus settings are changed, the transmit-
ter has to be reset before the new settings become ef-
fective.

The processor does not allow the following combinations:

no parity, 7 data bits, 1 stop bit: if this combination is given the
HMP230 program will change the number of stop bits to 2

even or odd parity, 8 data bits, 2 stop bits: if this combination is given
the program changes the number of stop bits to 1

Refer to the manuals of the PC and the terminal emulation program when
giving serial settings.

The RS 232C screw terminal cannot be used if an RS 485/422 serial module
or a current loop module is used. See Appendices 3 and 4 on how to install
and operate these modules.