Example – Vaisala HMP230 User Manual
Page 159

HMP230 SERIES
M210225en-B
Appendix 9: Re-gaining
153
Example:
>regen
Regeneration started, press any key to abort
>r
S
162 RH= 20.0 %RH T= 26.5 'C
heating is finished,
S
162 RH= 20.0 %RH T= 26.5 'C
settling time is on and
S
162 RH= 20.0 %RH T= 26.5 'C
outputs are locked
...
N
162 RH= 18.4 %RH T= 27.7 'C
settling time is over and
N
162 RH= 18.5 %RH T= 27.6 'C
outputs are released
N
162 RH= 18.8 %RH T= 27.5 'C
Re-gaining cannot be started if automatic re-gaining (heating/cooling) is on at
that moment, the temperature of the sensor exceeds 100 °C or the security lock
jumper is off. In these cases the 'Regeneration started...' text is not output by
the transmitter but the prompt '>' appears immediately.
Example:
>regen
>r
forced re-gaining failed
H
162 RH= 19.7 %RH T= 26.4 'C
as automatic re-gaining
H
162 RH= 19.7 %RH T= 26.4 'C
(heating) is going on
H
162 RH= 19.7 %RH T= 26.4 'C