Regen starting re-gaining manually, Example: activating status display, Deactivating status display – Vaisala HMP230 User Manual
Page 158

HMP230 SERIES
Appendix 9: Re-gaining
M210225en-B
152
Example: activating status display
>fst on
Form. status
:
ON
>r
N
0 RH= 20.8 %RH T= 25.6 'C
N = re-gaining off
N
0 RH= 20.7 %RH T= 25.6 'C
N
0 RH= 20.7 %RH T= 25.6 'C
N
0 RH= 20.8 %RH T= 25.6 'C
H
163 RH= 20.8 %RH T= 25.6 'C
H = heating on, 163
H
163 RH= 20.8 %RH T= 25.6 'C
describes the heating
H
163 RH= 20.8 %RH T= 25.6 'C
power
H
163 RH= 20.8 %RH T= 25.6 'C
S
163 RH= 20.8 %RH T= 25.6 'C
S = settling period on
S
163 RH= 20.8 %RH T= 25.6 'C
S
163 RH= 20.8 %RH T= 25.6 'C
S
163 RH= 20.8 %RH T= 25.6 'C
N
163 RH= 20.8 %RH T= 25.6 'C
N
163 RH= 15.0 %RH T= 31.5 'C
N
163 RH= 16.7 %RH T= 30.0 'C
N
163 RH= 17.4 %RH T= 29.1 'C
H
160 RH= 17.7 %RH T= 28.6 'C
H = new heating (new
H
160 RH= 17.7 %RH T= 28.6 'C
power)
deactivating status display
>fst off
Form. status
:
OFF
>r
RH= 20.1 %RH T= 26.4 'C
RH= 20.2 %RH T= 26.4 'C
RH= 20.1 %RH T= 26.4 'C
RH= 20.2 %RH T= 26.4 'C
REGEN
Starting re-gaining manually
REGEN
Re-gaining can also be started manually with command REGEN. The current
re-gaining parameters (duration and settling times stored with command CRI)
are used.
Example: starting re-gaining
>regen
Regeneration started, press any key to abort
>
The prompt '>' appears when the heating period is over. However, the
transmitter outputs are locked to values measured before re-gaining until the
settling time is over.