Cri changing re-gaining parameters, Fst displaying re-gaining status, Changing re-gaining parameters – Vaisala HMP230 User Manual
Page 157: Displaying re-gaining status

HMP230 SERIES
M210225en-B
Appendix 9: Re-gaining
151
CRI
Changing re-gaining parameters
Disconnect the security lock jumper!
CRI
With command CRI the values of the re-gaining parameters can be changed.
Example: changing the re-gaining interval to 480 min and the settling time
after sensor heating to 360 s
>cri
Interval
min :
720 ? 480
Duration
s :
120 ?
Temp diff
:
0.50 ?
Settl time
s :
240 ? 360
k
% :
95.00 ?
>
NOTE
To activate new parameter settings immediately,
command RESET must be given. Otherwise the
transmitter waits until the end of the old re-gaining
interval and then starts re-gaining using the new
parameters.
FST
Displaying re-gaining status
FST nnn
nnn = ON or OFF
With FST ON the status of the re-gaining is output at the beginning of the
output line.