HMP230 SERIES

M210225en-B

Appendix 9: Re-gaining

151

CRI

Changing re-gaining parameters

Disconnect the security lock jumper!

CRI

With command CRI the values of the re-gaining parameters can be changed.

Example: changing the re-gaining interval to 480 min and the settling time

after sensor heating to 360 s

>cri
Interval

min :

720 ? 480

Duration

s :

120 ?

Temp diff

:

0.50 ?

Settl time

s :

240 ? 360

k

% :

95.00 ?

>

NOTE

To activate new parameter settings immediately,
command RESET must be given. Otherwise the
transmitter waits until the end of the old re-gaining
interval and then starts re-gaining using the new
parameters.

FST

Displaying re-gaining status

FST nnn

nnn = ON or OFF

With FST ON the status of the re-gaining is output at the beginning of the
output line.