Vaisala HMP230 User Manual
Page 56

HMP230 SERIES
User's Guide
M210225en-B
50
6.4.2.6. Selecting the output units
UNIT x
x =
m(etric units)
n(on-metric units)
quantity
metric
non-metric
RH = relative humidity
%RH
%RH
T = temperature
°C
°F
Td = dewpoint temperature
°C
°F
a = absolute humidity
g/m
3
gr/ft
3
x = mixing ratio
g/kg
gr/lb
h = enthalpy
kJ/kg
Btu/lb
Tw = wet bulb temperature
°C
°F
6.4.2.7. Setting the averaging time
Disconnect the security lock jumper!
FILT nnnn
nnn
=
averaging time (0 - 1024 seconds)
This command is used to set and inspect the averaging time during which the
individual measurement samples are integrated to get an averaged reading.
The time can be set in seconds within the range of 0 - 1024 (0 = no averaging
time).
For example:
>FILT
Filter (S):
0 ? 1024
>
>FILT 100
Filter (S):
100
>
6.4.2.8. Setting the transmitter address
ADDR aa
aa =
address (0...99)
Example: transmitter is given address 99
>ADDR
Address : 2 ? 99