HMP230 SERIES
User's Guide

M210225en-B

50

6.4.2.6. Selecting the output units

UNIT x

x =

m(etric units)
n(on-metric units)

quantity

metric

non-metric

RH = relative humidity

%RH

%RH

T = temperature

°C

°F

Td = dewpoint temperature

°C

°F

a = absolute humidity

g/m

3

gr/ft

3

x = mixing ratio

g/kg

gr/lb

h = enthalpy

kJ/kg

Btu/lb

Tw = wet bulb temperature

°C

°F

6.4.2.7. Setting the averaging time

Disconnect the security lock jumper!

FILT nnnn

nnn

=

averaging time (0 - 1024 seconds)

This command is used to set and inspect the averaging time during which the
individual measurement samples are integrated to get an averaged reading.
The time can be set in seconds within the range of 0 - 1024 (0 = no averaging
time).
For example:

>FILT
Filter (S):

0 ? 1024

>

>FILT 100
Filter (S):

100

>

6.4.2.8. Setting the transmitter address

ADDR aa

aa =

address (0...99)

Example: transmitter is given address 99

>ADDR
Address : 2 ? 99