Serial commands, Reg re-gaining settings, Cr re-gaining parameters – Vaisala HMP230 User Manual
Page 156: Serial, Commands, Re-gaining settings, Re-gaining parameters

HMP230 SERIES
Appendix 9: Re-gaining
M210225en-B
150
5.
SERIAL COMMANDS
REG
Re-gaining settings
REG
REG nnn
nnn = ON or OFF
Command REG is used to check the re-gaining parameter settings and to
activate or inactivate the automatic re-gaining.
Examples: checking the current settings
>reg
REG OFF
automatic re-gaining is inactive
Interval
min :
720
Duration
s :
120
Temp diff
:
0.50
Settl time
s :
240
k
% :
95.00
>
activating the automatic re-gaining
>reg on
REG ON
>
deactivating the automatic re-gaining
>reg off
REG OFF
>
CR
Re-gaining parameters
CR
Command CR can be used to check the values of re-gaining parameters.
Example:
>cr
Interval
min :
720
Duration
s :
120
Temp diff
:
0.50
Settl time
s :
240
k
% :
95.00
>
See also other documents in the category Vaisala Humidifiers:
- Calibration of Digital Transmitters with HMI41 (36 pages)
- Calibration of Series HMDW2030 and HMP130 Transmitter with HMI41 (14 pages)
- Calibration of Series HMDW6070 and HMP140 Transmitter with HMI41 (30 pages)
- HM34 (30 pages)
- HM40 (47 pages)
- HM44 (52 pages)
- HM70 (83 pages)
- HMD40 (1 page)
- HMD60 (4 pages)
- HMD70 (18 pages)
- HMDW110 (62 pages)
- HMDW80 (51 pages)
- HMI41 (74 pages)
- HMP41 (72 pages)
- HMK15 (39 pages)
- HMM100 (71 pages)
- HMM105 (23 pages)
- HMM211 (42 pages)
- HMM212 (36 pages)
- HMM213 (52 pages)
- HMP140 (28 pages)
- HMP155 (84 pages)
- HMP228 (115 pages)
- HMP240 (130 pages)
- HMP260 (118 pages)
- HMP60 (71 pages)
- HMT100 (52 pages)
- HMT120 (87 pages)
- HMT130 (95 pages)
- HMT140 (76 pages)
- HMT310 (88 pages)
- HMT310 (105 pages)
- HMT330 (209 pages)
- HMT360 (97 pages)
- HMT360 (63 pages)
- HMT360N (110 pages)
- HMW40 (1 page)
- HMW90 (110 pages)
- SHM40 (68 pages)
- RDP100 (14 pages)