200
Chapter 10
Troubleshooting
Error Messages
Error Messages
If an error condition occurs in the signal generator, it is reported to both the front panel display error queue
and the SCPI (remote interface) error queue. These two queues are viewed and managed separately; for
information on the SCPI error queue, refer to the Programming Guide.
NOTE
When there is an unviewed message in the front panel error queue, the
ERR
annunciator
appears on the signal generator’s display.
Error Message File
A complete list of error messages is provided in the file errormesages.pdf, on the CDROM supplied with
your instrument.
In the error message list, an explanation is generally included with each error to further clarify its meaning.
The error messages are listed numerically. In cases where there are multiple listings for the same error
number, the messages are in alphabetical order.
Characteristic
Front Panel Display Error Queue
Capacity (#errors)
30
Overflow Handling
Circular (rotating).
Drops oldest error as new error comes in.
Viewing Entries
Press:
Utility
>
Error Info
>
View Next
(or
Previous
)
Error Message
Clearing the Queue
Press:
Utility
>
Error Info
>
Clear Error Queue(s)
Unresolved Errors
a
a. Errors that must be resolved. For example, unlock.
Re-reported after queue is cleared.
No Errors
When the queue is empty (every error in the queue has been read, or the queue is cleared), the
following message appears in the queue:
0
No Error Message(s) in Queue