Agilent Technologies E8247C PSG CW User Manual
Page 134
![background image](/manuals/37799/134/background.png)
124
Chapter 6
Custom Arb Waveform Generator
Working with User-Defined Setups (Modes)-Custom Arb Only
Recalling a User-Defined Custom Digital Modulation State
In this procedure, you learn how to select (recall) a previously stored custom digital modulation state from
the Memory Catalog (the
Catalog of DMOD Files
).
1. Press
Preset
.
2. Press
Mode
>
Custom
>
ARB Waveform Generator
>
Setup Select
.
3. Press
More (1 of 2)
>
Custom Digital Mod State
.
4. Press
Select File
to select a custom modulation state from the
Catalog of DMOD Files
.
The user-defined custom digital modulation state should now be recalled from non-volatile memory.
Because the RF output amplitude, frequency, and operating state settings are not stored as part of a
user-defined digital modulation state file, they must still be set or recalled separately. For more
information, refer to