Signal Generator Locks Up
3. Release the
Preset
key.
4. To continue with the sequence, press
Continue
(to abort with no lost files, press
Abort
).
5. When the sequence concludes:
a. Cycle power.
Cycling power restores all previously installed options. Because calibration files are restored from
EEPROM, you should see several error messages.
b. Perform the DCFM/DC
Φ
M calibration.
Refer to the
DCFM/DC
Φ
M Cal
softkey description in the Key Reference.
c. Agilent Technologies is interested in the circumstances that made it necessary for you to initiate this
procedure. Please contact us at the appropriate telephone number listed in
We would like to help you eliminate any repeat occurrences.