Rf output power problems, Rf output power too low, The power supply has shut down – Agilent Technologies E8247C PSG CW User Manual
Page 199
![background image](/manuals/37799/199/background.png)
Chapter 10
189
Troubleshooting
RF Output Power Problems
RF Output Power Problems
Check the
RF ON/OFF
annunciator on the display. If it reads
RF OFF
, press
RF On/Off
to toggle the RF
output on.
RF Output Power too Low
1. Look for an
OFFS
or
REF
indicator in the
AMPLITUDE
area of the display.
OFFS
tells you that an amplitude offset has been set. An amplitude offset changes the value shown in the
AMPLITUDE
area of the display but does not affect the output power. The amplitude displayed is equal to
the current power output by the signal generator hardware plus the value for the offset.
To eliminate the offset, press the following keys:
Amplitude
>
More (1 of 2)
>
Ampl Offset
>
0
>
dB
.
REF
tells you that the amplitude reference mode is activated. When this mode is on, the displayed
amplitude value is not the output power level. It is the current power output by the signal generator
hardware minus the reference value set by the
Ampl Ref Set
softkey.
To exit the reference mode, follow these steps:
a. Press
Amplitude
>
More (1 of 2)
.
b. Press
Ampl Ref Off On
until Off is highlighted.
You can then reset the output power to the desired level.
2. If you are using the signal generator with an external mixer, see
“Signal Loss While Working with a
.
3. If you are using the signal generator with a spectrum analyzer, see
“Signal Loss While Working with a
Spectrum Analyzer” on page 192
.
The Power Supply has Shut Down
If the power supply is not working, it requires repair or replacement. There is no user-replaceable power
supply fuse. Refer to the Service Guide for instructions.