Configuring hardware – Agilent Technologies E8247C PSG CW User Manual
Page 153
Chapter 6
143
Custom Arb Waveform Generator
Configuring Hardware
Configuring Hardware
•
see
To Set a Delayed, Positive Polarity, External Single Trigger
Using this procedure, you learn how to utilize an external function generator to apply a delayed
single-trigger to a custom multicarrier waveform.
1. Connect an Agilent 33120A function generator or equivalent to the signal generator PATT TRIGGER IN
port, as shown in
Figure 6-1
2. On the signal generator, press
Preset
.
3. Press
Mode
>
Custom
>
Arb Waveform Generator
.
4. Press
Multicarrier Off On
until On is highlighted.
5. Press
Trigger
>
Single
.
6. Press
Trigger
>
Trigger Setup
>
Trigger Source
>
Ext
.
7. Press
Ext Polarity Neg Pos
until Pos is highlighted.
8. Press
Ext Delay Off On
until On is highlighted.
9. Press
Ext Delay Time
>
100
>
msec
.
The Custom Arb Waveform Generator has been configured to play a single multicarrier waveform
100 milliseconds after it detects a change in TTL state from low to high at the PATT TRIG IN rear panel
connector.
10. Set the function generator waveform to a 0.1 Hz square wave at an output level of 0 to 5V.