beautypg.com

Diagram 22, Diagram 23, 6 view a reference setup – Setra System Model 869 User Manual

Page 25: Diagram 16

background image

32

Electrical Lines

Unit Under Test

(Model 260)

Pneumatic Lines

Diagram 22

_

+

EXC

24VDC

SIG

(+)

SIG

(-)

COM

Current Internal

Excitation

EXC COM SIG

EXC

24VDC

SIG

(+)

SIG

(-)

COM

3-Wire Internal

Excitation

EXC

24VDC

SIG

(+)

SIG

(-)

COM

4-Wire Internal

Excitation

EXC COM +SIG -SIG

_

+

EXC

24VDC

SIG

(+)

SIG

(-)

COM

_

Current External

Excitation

+

External

Power

Supply

EXC COM SIG

EXC

24VDC

SIG

(+)

SIG

(-)

COM

3-Wire External

Excitation

_

+

External

Power

Supply

EXC

24VDC

SIG

(+)

SIG

(-)

COM

4-Wire External

Excitation

EXC COM +SIG -SIG

_

+

External

Power

Supply

Diagram 23

EXC

24VDC

SIG

(+)

SIG

(-)

COM

P-Switch SPST

Unit Under Test

(Model 269)

EPIC Cable

EPIC Cable Plug
(Calibrator End)

EPIC Cable Plug

(UUT End)

Standard 869

869 Expert System

25

6.6 VIEW A REFERENCE SETUP

Screen Instructions (Diagram 16):
1. Tap the REF tab to view the different reference data.
2. Tap on REF 1 or REF 2 radio button.

Reference Setup - The Model 869 is shipped with either one or two inter-
nal reference transducers and are referred to as REF 1 or REF 2 respec-
tively. REF 1 is the lowest range and REF 2 is the highest range. If only
one reference is installed, it will be located in REF 1.


REF ID - the serial number of the reference unit.

Input/Output Parameters - This section tells you the pressure range and
the output of the reference transducer.

Tare Button - Allows the user to “zero” the reference unit.

Reference Location - Shows the high and low ranges of the reference
unit, (the Model 869 automatically selects the proper reference sensor
according to the full scale range of the presently selected UUT).

Diagram 16

Select Reference

Number

Select Reference

Tab