Setra Systems, Inc.
Adding a Site Offset to SDI12 Output
In some applications it is necessary to measure a site value including some
reference data point, for example sea level. Assuming the sensor is installed 10m
above sea level, then it is possible to have all reported readings add 10m, thus
acquiring the data without the need for further modification. To add a site offset
then:
1. Determine the offset to be applied.
2. Add the offset to the pressure units (e.g. aXU1,10!)
3. Save new calibration to nonvolatile memory (e.g. aXS0!)
4. Verify sensor now measures 10m for 0 applied pressure.
10. WARRANTY
The company warrants its products to be free from defects in material and
workmanship in normal use and service for a period of two years from date of
shipment. The Company reserves the right and option to repair or replaceupon
evaluation of the returned original part. Modification, misuse, attempted repair by
others, improper installation or operation shall render this guarantee null and
void. The Company makes no warranty of merchantability or fitness for a part or
purpose.
11. Servicing
The transducer cannot be repaired locally and if damaged should be returned to
ourselves at the address shown below when a replacement/repair is required.
Setra Systems, Inc.
159 Swanson Road
Boxborough, MA 01719
Attn: Repair Department
12. Return to Factory
PLEASE NOTE: All returns must be accompanied with Setra’s “Calibration and
Service Order” return form. The form can be downloaded at:
http://www.setra.com/tra/repairs/pdf/webrepair.pdf
Any instrument returned without Setra’s authorized form completely filled out
will be quarantined and no action will occur. It may ultimately be returned to you
and subject to a transportation charge.
159 Swanson Road, Boxborough, MA 01719/
Toll Free: 800-257-3872/ Phone: 978-263-1400/
Fax: 978-264-0292/Email: [email protected]/Web: www.setra.com
ISO
✓
Certified
9001:
2000
Document Outline
- Contents
- 1.0 Introduction
- 2.0 Safety Instructions
- 3.0 Hazardous Products
- 4.0 GENERAL
- 5.0 Action on Receipt
- 6.0 Installation
- 6.1 Lowering the Pressure Probe
- 6.2 Electrical Installation
- 7.0 Operation
- 7.1 Analogue Output Probes
- 7.2 SDI12 Operation
- 8. Maintenance
- 8.1 Self diagnostics
- 8.2 Calibration Adjustment
- 8.3 Cleaning
- 9. Advanced Operation
- 10. WARRANTY
- 11. Servicing
- 12. Return to Factory