Apply power to calibrator with

AC adapter or charge battery

.

T

urn on power switch located in the top right corner

. Power is on when green light illuminates - the pocket PC will

ash an orange light.

Push Pocket PC power button in top right of the unit. It will display a “start” screen.

Tap “start” and select Setra Calibrator icon from “last

programs run” or “programs” fi

le from drop down menu.

Calibrator starts in “Real

T

ime” mode.

Tap UUT

(Unit Under

Test) tab and select a stored UUT

ID that corresponds to the pressure sensor

you wish to test - (tap the down arrow in the top box to view all of the stored UUT

test profiles). Power is supplied to test pressure sensor

when UUT

ID is selected. If a needed ID is not found, see Creating and Modifying UUT

ID’

s below

.

Plug in pressure tubing and electrical cables into marked sockets on the top panel. Connect pressure tubing to “high” and “low”

pressure

fittings. Hook up connections as follows:

1. 4-20mA

current loop operation, connect +EXC to red 869 banana jack and -EXC to black jack.

2. For 3-wire voltage output operation, connect +EXC to red electrical jack, -EXC common to black jack, and +Output to green

jack.

Return to “Real

T

ime” screen to verify correct zero pressure output of test pressure sensor

. T

ap “tare” button to re-zero onboard pressure

references (approx 10 seconds).

Y

ou can also check full-scale output by tapping “+FS” (+ full scale) button.

Y

ou can also make zero and

span adjustments to test pressure sensor at this time.

Go to “T

est” screen and tap “Do

Test” button to initiate automatic calibration test.

The calibrator will apply test pressures and record output

results until “test complete” fl

ag is displayed.

Y

ou can save the test data by tapping the “Save

Test Results” button.

The data is saved with

Micro-Cal

®

Model 869 Quick Start Guide

To create a new UUT

ID go to the

UUT

screen and tap “

new

” button. Next, type in a new

UUT ID

in the upper box. Enter data by tapping

the keyboard icon in lower right of screen. Characters are entered on screen at the cursor when mini-keyboard characters are ta

pped.

An

existing ID template must be selected in “

copy from

” box. It is suggested that an existing ID template be chosen that closely matches the

pressure sensor desired to test either in pressure range or output type.

Tap “

ok

”. When screen returns to UUT

, any ID values can be modi

-

fied using keypad. Use dropdown boxes to change excitation, outpu

t type, or number of test points. Hit “save” button and new tes

t profi

le is

ready to use.

To modify a UUT

ID, select desired UUT

ID from dropdown screen, make changes, and tap “

save

”.

The newly created or modifi

ed UUT

IDs

will be available from dropdown UUT

ID screen.

Creating and Modifying UUT ID’

s

I

II

III

IV

V

VI

VII

Setra Systems, Inc., 159 Swanson Rd, Boxborough, MA, 01719

User Manual

Micro-Cal™

Model 869

Very Low Pressure Generating and Documenting Calibrator

800.257.3872

• Toll Free

978.264.0292

• Fax

setra.com

• Web Site

pressure transducers

Software Version

V3.02

SS869 2/03/12 Rev

. D

Software Version V3.21