Mpcal= (multipoint calibration) – PNI TCM 2.6 User Manual
Page 30

PNI Sensor Corporation
DOC#1009269 r11
TCM2.5 & TCM2.6 User Manual
Page 30 of 49
mpcal=
(Multipoint Calibration)
Description
Initiates a multipoint calibration. This command is only valid when used in
conjunction with Enter Continuous Sampling Mode (go). The TCM2.5/2.6 samples its
sensors and adds the data point to the current set of calibration data.
Syntax
mpcal=e
:
go
.......
...... take data in continuous mode
h
:
mpcal=d
Valid Values
Input
Description
e
e, enable multipoint calibration
d
d, disable multipoint calibration (default setting)
Table 8: Valid
mpcal= Values
Output
HnVnMn.nn
The TCM2.5/2.6 only reports a calibration score after it has been put in Standby mode (h), and
multipoint calibration has been disabled (mpcal=d)
Example
mpcal=e
:
go
........
....... take data in continuos mode
h
:
mpcal=d
H4V3M5.00:
Query Response
mpcal?
:mpcal=
Related Commands
lc? (Query Last Calibration Score)