beautypg.com

Prevent malfunction – Dwyer TIC User Manual

Page 10

background image

9

Prevent Malfunction

Read on to learn about preventing malfunction of IR camera

Avoid damaging the detector of the IR camera

Warning:

Do not aim the IR camera directly into the sun or at other

intense heat source which could damage the detector of the IR

camera.

Avoid Condensation Related Problems

Moving the IR camera rapidly between hot and cold temperatures may

cause condensation (water droplets) on its external and internal

surfaces.

You can avoid this by placing the IR camera in the plastic case (bundle)

and letting it adjust to temperature changes slowly before removing it

from the case.

If Condensation Forms Inside the IR Camera

Stop using the camera immediately if you detect condensation.

Continue to use may damage the IR camera. Remove the PC card, and

battery or a household power source, from the IR camera and wait until

moisture evaporates completely before resuming use.

Extended Storage

When not using the IR camera for extended periods of time, remove the

battery from the IR camera or battery charger and store the IR camera in

a safe place. Storing the IR camera for extended periods with battery

installed will run down the battery.

Different Models

Some advanced functions are included only in TIC-30 and TIC-20,

TIC-10 only has the fundamental function

Right Reserved

Dwyer reserves the right to change the functions and configurations of

our products without prior notice.