Appendix b — default settings – Metex 8760CLP Total Free Chlorine & pH Analyzer User Manual
Page 62
![background image](/manuals/369857/62/background.png)
AQUAMETRIX INC.
1-800-742-1413 www.aquamerix.com
62
APPENDIX B — Default Settings
The following program settings are the default settings for the analyzer. New analyzers will have these
settings unless the setup has already been customized for your application.
Outputs
Output 1 Output 2
Input to be transmitted
tFCl
°C
Low setting
0.00
0.0
High setting
2.00
100.0
ON/OFF switch
ON
ON
Alarms
Alarm A
Alarm B
Input for alarm
tFCl
tFCl
Alarm function
HI
LO
ON/OFF switch
ON
ON
Set-point
0.60 ppm
0.20 ppm
Differential
0.10 ppm
0.10 ppm
Unit
1E-6 (ppm) 1E-6 (ppm)
Delay
0 s
0 s
Global Units
Metric units: temperature in degrees Celsius, chlorine concentration in parts per million (ppm)
Alarm Contacts
Configured normally open
Security
Not enabled
pH and Temperature Compensation Method for Chlorine
Automatic TC using temperature input; automatic pH compensation using pH input
Membrane Compensation
Enabled
Timer Feature
Not enabled