beautypg.com

Specifications – Metex 8760CLP Total Free Chlorine & pH Analyzer User Manual

Page 11

background image

AQUAMETRIX INC.

1-800-742-1413 www.aquametrix.com

11

Specifications

AM-A2104034 Chlorine Sensor

Measurement Range
Free Available Chlorine (HOCl)....................................................................0.00 mg/L to 2.00 mg/L

Total Free Available Chlorine (HOCl + OCl⎯)................................................0.00 mg/L to 5.00 mg/L


Minimum Temperature

.................................................................................................0 °C (32 °F)

Maximum Temperature

............................................................................................80 °C (176 °F)


Maximum Pressure

..............................................................................................621 kPa (90 psi)


Principle of Operation

.......................................................................................................Galvanic

Electrode Materials

Cathode......................................................................................................................................Gold

Anode.......................................................................................................................................Silver

Minimum Flow Velocity

........................................................................................15 cm/s (0.5 ft/s)


Wetted Materials

..............................................................................................CPVC, PTFE, Viton


Electrode Dimensions

Diameter.....................................................................................................................2.3 cm (0.9 in)

Length.......................................................................................................................16.5 cm (6.5 in)

Process Connection

..............................................................fixed in the flow cell via a 1 in MNPT

CPVC quick connect insertion fitting


Sensor Cable

...............................................................................4 conductor; 1.5 m (5 ft) in length

with 5-pin DIN connector


Weight

.........................................................................................................................0.5 kg (1.1 lb)


Shipping Weight

........................................................................................................0.9 kg (2.0 lb)


Shipping Dimensions

.................................................................................30 cm

× 23 cm × 23 cm

(12 in

× 9 in × 9 in)